Troubleshooting:
Before troubleshooting please take these words of "Balance Beam Wisdom":
Balance Beams are a very old and basic design of scale. They date back to
the ancient Egyptian Empire and Balance Beam scales were used to build the
pyramids. Thus, the design is quite simple - a cup (hidden on modern
beam scales) filled with metal loose weights and some metal sliding weights.
Therefore to fix a Balance Beam scale requires very basic steps - such as
adding/removing weight from the cup or using a wrench to make the beam operate
smoothly (if it's touching or rubbing as it operates). First, before
troubleshooting be certain that you have correctly unpacked the scale and
removed all rubber stoppers (you'd be surprised how many times this happens).
Also be sure that the sliding weights are all at their proper ZERO position.
Please be 100% sure that you have removed both rubber stoppers and all packing
bands/materials.
1) "My Scale Won't Zero": Sometimes in transport the beam will
somehow become unbalanced. This means adjusting the zero knob doesn't make the
scale zero properly. If this happens to your scale please know that you
can manually add or remove weight to the "Balance Cup". The Balance cup is
located underneath the round metal tray. Simply spin the tray off
(counterclockwise) to access the Balance Cup and add or remove weight as
follows:
If your scale would not zero because it was reading too light (meaning you
would have had to press down on the tray to make it zero), then add a
small amount of weight to the cup (just a coin or paperclip, you can add
or remove weight as you require).
If your scale would not zero because it was reading too heavy (meaning you
would have had to pull up on the tray to make it zero), then remove a
small amount of weight from the cup (just remove a tiny amount, you can
remove or add weight as you require).
If you are trying to compensate for a small bowl or tray, you can
remove weight from the cup until the scale Zeros. The Balance Beam
is a very basic scale and if it Zeros & moves freely, then it always reads
accurately.
2) "The Beam is Rubbing / Touching": In order for this scale to
work properly, it must operate smoothly without touching or rubbing.
Sometimes in transport the beam will get shifted and touch the internal magnets
by the front of the beam (where the "0" indicator is). If this
happens please try to manually adjust the beam using a wrench on the main bolt
under the scale. Simply twist the beam to make it operate smoothly.
If you cannot make the beam operate smoothly then the scale will need to be sent
in for service.
3) "My Scale won't go back to Zero": A triple beam works with a
set of sliding weights. If these weights are not put back EXACTLY to their
Zero position, the scale won't return to zero. Be sure that
when you Zero the scale you note where you have put all 3 weights exactly,
then when you are done weighing be sure they are in the exact same place.
Based on the laws of physics it would be impossible for the beam not to return
to Zero if the weights are put back to the same spot (unless the beam is
rubbing/touching or if the surface you are weighing on is not level).
Construction and Application
The My Weigh MB-2610 Single-Pan Balance is of unequal arm type with three beams.
The beam is made of high quality aluminum alloy and the plane is of agate which
can improve accuracy of use and lengthen life of operation. To speed up weight
readings, the magnetic damping system brings the pointer to rest with a minimum
number of swings.
It is widely suitable for
physical-chemistry experiments and analytical weighing in chemical works,
medical and hygiene, food-stuff, agriculture, textiles, electronics, mines,
scientific research institutions, universities and colleges, etc.
-
Unpacking:
-
Carefully remove the balance
and the separate poise (sliding weight) from the
-
protective carton. You will
find a slit rubber washer lodged underneath the
-
platform, and one rubber
washer located above the pointer. The washers are to
-
be removed from the scale.
Set-up:
After placing the balance on a
smooth, flat surface, slide each separate poise (sliding weight) into it’s zero
slot. With all poises at the zero position, the pointer should be near zero.
Carrying the Balance:
Be sure all riders are back
to the zero point. Place one hand under the balance and the other hand on the
support (arm) to carry the balance.
Zeroing:
For exact zero, adjust the
knurled knob which is located at the left end of the beam. It is advisable to
check the zero adjustment periodically.
Weighing Process:
Before Weighing:
Zero
the balance before you determine the weight of any substance.
a) Slide all of the riders back to the zero point.
b) Check to see that the pointer swings freely along the scale.
c) Use the adjustment knob to obtain an equal swing of the beams, if
necessary. You do not have to wait for the pointer to stop at the zero point.
The beam should swing an equal distance above and below the zero point. You must
repeat this procedure to "zero" the balance every time you use it.
-
Begin Weighing:
-
Place the specimen on the
center of the platform and proceed as follows:
-
1.
Starting with the largest capacity beam (500 g), move
the 500 g poise to the
-
right to the first notch which
causes the pointer to drop, then, move it back one
-
notch, causing the pointer to
rise.
-
2. Repeat procedure with the 100 g poise.
-
3.
Slide the 10 g poise to the position that brings the
pointer to rest at zero.
-
The weight of the specimen is
the sum of the values of all poise positions, read
-
directly from the graduated
beams.
NOTES:
-
A)
Never put a hot object directly on the balance pan. Any dry chemical
that is to be massed should be placed on paper or in a container.
-
B)
Never pour
chemicals directly on the balance pan. Remember to weigh or zero the paper
or container before adding the substance.
-
C)
Remember; the
pointer does not have to stop swinging, but the swing should be an equal
distance above and below the zero point on the scale.
Attachment
weights:
Total capacity is either 2610
grams or 5 pounds, 2 ounces when attachment weights are suspended from the
pivots. Without the weights, the capacity is
610 grams (1 pound, 2 ounces).
Care and maintenance:
Keep the balance clean at all
times. Most dirt or foreign matter can easily be
removed with an air syringe, but
a piece of adhesive-backed tape pressed against the magnet faces will keep them
free from dirt. Never apply lubricants to the knives or bearings and don’t
allow dirt or foreign materials to accumulate on any part of the beam. The Beam
must be kept clean in order to provide years of accurate stable readings.
Specifications
Capacity:
Without attachment weights: 610g
With included attachment weights: 2610g
Readability: 0.1g
Weighing Units: grams
Beam Calibrations: 10 x 0.1g, 100 x 10g, 500 x 100g
Features:
Three
beam direct reading
Attachment Weights Included
Magnetic Damping System
Troubleshooting:
Before
troubleshooting please take these words of "Balance Beam Wisdom": Balance
Beams are a very old and basic design of scale. They date back to the ancient
Egyptian Empire and Balance Beam scales were used to build the pyramids. Thus,
the design is quite simple - a cup (hidden on modern beam scales) filled with
metal loose weights and some metal sliding weights. Therefore to fix a Balance
Beam scale requires very basic steps - such as adding/removing weight from the
cup or using a wrench to make the beam operate smoothly (if it's touching or
rubbing as it operates). First, before troubleshooting be certain that you
have correctly unpacked the scale and removed all rubber stoppers (you'd be
surprised how many times this happens). Also be sure that the sliding weights
are all at their proper ZERO position. Please be 100% sure that you have
removed both rubber stoppers and all packing bands/materials.
1)
"My Scale Won't Zero": Sometimes in transport the beam will somehow become
unbalanced. This means adjusting the zero knob doesn't make the scale zero
properly. If this happens to your scale please know that you can manually add
or remove weight to the "Balance Cup". The Balance cup is located underneath
the round metal tray. Simply spin the tray off (counterclockwise) to access
the Balance Cup and add or remove weight as follows
If your
scale would not zero because it was reading too light (meaning you would have
had to press down on the tray to make it zero), then add a small amount of
weight to the cup (just a coin or paperclip, you can add or remove weight as
you require).
If your
scale would not zero because it was reading too heavy (meaning you would have
had to pull up on the tray to make it zero), then remove a small amount of
weight from the cup (just remove a tiny amount, you can remove or add weight as
you require).
If you
are trying to compensate for a small bowl or tray, you can remove weight from
the cup until the scale Zeros. The Balance Beam is a very basic scale and if
it Zeros & moves freely, then it always reads accurately.
2)
"The Beam is Rubbing / Touching": In order for this scale to work properly,
it must operate smoothly without touching or rubbing. Sometimes in transport
the beam will get shifted and touch the internal magnets by the front of the
beam (where the "0" indicator is). If this happens please try to manually
adjust the beam using a wrench on the main bolt under the scale. Simply twist
the beam to make it operate smoothly. If you cannot make the beam operate
smoothly then the scale will need to be sent in for service. |