my weigh scales logo scales background scale circle
HomePocket and Mini ScalesMedium ScalesLarge ScalesOnline StoresSupportContact Us
 
 
FAQ
General Tech Support
 
 
in alphabetical order:
200-Z Manual
1scale Manual and Calibration guide
3001P and 7001DX Series Manual
AXE Manual
BodyFlight Manual
C200 Convertible Manual
Durascale Manual
Flipscale Classic Manual
Flipscale Classic Advanced Programming
Flipscale Interactive Manual
Flipscale F2 Manual
GemPro  manual and advanced guide
HD-150/300 Manual
iScale Manual
i250 Manual
i500, i1200 & i2600 Manuals
i5000 & i2500 Manual
iBalance 201 Manual
iBalance 101 Manual
KD-600 Manual
LS-30 Manual
MBS-2010 Babyscale Manual & Calibration
MX-50 Manual
MX-300 & MX-120 Manual
Palmscale (all) Manual and Calibration Guide
OMNI Manual
Palmscale 1.0 and 2.0 Advanced Guide and Repair
Palmscale 3.0 and 4.0 Advanced Guide and Repair
Palmscale 5.0 Advanced Programming and Repair Guide
Phoenix Talking Scale Manual
Phoenix Body Fat Scale Manual
Phoenix Body Fat Scale
Common Questions
Pointscale Manual
Proscale Manual & troubleshooting
Tempest manual
Touchscale Manual
Touchscale Advanced Programming
Triton Manual
Triple Beam Manual & Troubleshooting
UltraShip Manual
XL440 Manual
 
AXE2 Owner's Manual & Troubleshooting

Low Batteries & bad battery connections are the #1 cause of scale malfunction and inaccuracy!   We test all of our scale returns from consumers. Fully 60% of consumer returns are battery related problems. This sounds silly but it's true! A scale will perform slowly, or read inaccurately when it has low batteries. Please replace the batteries often (and only use good quality batteries). We include good quality batteries with all of our scales but batteries can run low in storage.  If your scale simply won't turn on while on Battery power,  this is often caused by loose battery connections.  Battery prongs (terminals) are made of metal. They must be making good contact with your batteries in order for the scale to power on.  You can use a paperclip to slightly bend the battery prongs to make them have a better connection. Also some poorly designed batteries have recessed or partially obstructed battery terminals. This may cause your prongs to be touching the plastic housing of the battery instead of the metal battery terminal.  

AXE2 TROUBLESHOOTING & OPERATION NOTES:

1) If the Display ever becomes locked showing "Out2", please recalibrate the scale as outlined in the manual

2) If the Display ever becomes locked on ErrE, LLLL or EEEE, this indicates that the scale was shocked, dropped or otherwise damaged and the delicate weighing sensors have been damaged. You can try recalibrating the scale (If the sensor has not been hurt too badly it will work again after recalibration). Otherwise you will have to follow the warranty instructions that came with your scale.

3) If the display becomes locked on 8888, this often indicates low batteries. However sometimes it also may indicate a serious zero mark error. This means when you turn the scale on, it can't determine what zero is (a slight zero mark error will cause situation #1 above) Thus, if new batteries do not fix this error the scale will have to be sent to us for replacement under our warranty program.

4) If the display shows UNST, this means the scale is not stable. Please try operating the scale on a more stable surface and be sure nothing is on the tray or stuck under the tray when you turn it on. If this situation persists, it may be an indication of fatal load cell damage. Try calibrating the scale - if this does not work please follow the warranty instructions.

Although the AXE is designed to be extremely durable, It's important that you never overload or drop/shock the scale. Scales are delicate instruments and unlike Cellular phones, scales have delicate sensors that determine how much an item weighs. If you drop or shock your scale, these sensors "feel" the shock and are sometimes destroyed. This happens with all digital scales. We design our scales to be as resistant to shock or drops as possible, however there is no way for us to protect 100% against load cell or sensor damage. A welltreated scale will provide years of reliable and accurate weighing. However an abused scale will only work until it's sensors are damaged.

 

OPERATION_____________________________________________

Battery Replacement:

1) Three #357 button batteries are required.

2) To install batteries:

a) Release the battery cover by sliding out-wards.

b) Place batteries into battery compartment aligned correctly.

c) Replace battery cover.

DO NOT USE EXCESSIVE FORCE & DO NOT PRESS ON THE TRAY!!!

3) The scale is now ready for battery operation.

(Advanced users only) CALIBRATION______________________

IMPORTANT: This scale was professionally factory calibrated before shipment and does NOT need to be recalibrated by the end user. However, if you wish to recalibrate your scale we provide these instructions for calibration: Repeat calibration if the scale ever shows Out2. Incorrect calibration can occur if you do not follow the steps exactly. If your scale does not perform accurately, please try replacing your batteries before you calibrate. You will need a 200gram weight to calibrate this scale. NOTE: if you do not have access to a 200g weight you can purchase one at your local store or in emergency situations you can use 200g of coins or weights (ie 40 US Nickels = 200g) as a 200g weight..

The AXE2-300 requires a 200gram weight to calibrate

The AXE2-600 requires a 500gram weight to calibrate

1) Place the scale on a flat, vert stable surface

(Be certain the scale is OFF)

2) Press the key, the display will show "HELLO" and then press and hold [UNIT] key until the display shows CAL, then release the key and the value will show the A/D value

3) Press the key again, the display will show 2ERO and then

200.0 or 500.0

4) Gently place weight(s) on the scale and wait for a few seconds, the display will show PASS, then return to calibration value.

5) Now calibration is complete.
1. Press key. Turn on the power again, place the weight(s) on the platform to make sure whether it's correct, if not repeat steps 2 - 5.

2. Return to weighing mode
Press for 3 seconds to turn off the power. Press again to turn on the power and the scale return to weighing mode.

NOTE: If after calibration your scale does not read accurately, this indicates calibration error and the calibration process should be repeated more slowly.

Please calibrate on a very stable flat surface

 

INACCURACY / ERROR___________________________________

*Switch on When turn on the scale, the display will show "HELLO" and then reset to 0.

* Overload When an applied load exceeds the capacity. "EEEE" will appear on the display. Remove the excessive load immediately! !Remember: You can permanently damage the scale by overloading it!

* Negative Value Any tarred value or a value left in memory will be displayed as a negative number once all weight is removed from the unit. Press tare to re-Zero the unit.

* Auto off An auto shut off feature is provided to conserve battery power. The unit will automatically turn off after approximately 120 seconds of inactivity.

*Stable Indicator There is a stable indicator that appears on the left side of the LCD whenever the scale is stable and at Zero

1) If the Display ever becomes locked showing "Out2", please recalibrate the scale as outlined in the manual

2) If the Display ever becomes locked on ErrE, LLLL or EEEE, this indicates that the scale was shocked, dropped or otherwise damaged and the delicate weighing sensors have been damaged. You can try recalibrating the scale (If the sensor has not been hurt too badly it will work again after recalibration). Otherwise you will have to follow the warranty instructions that came with your scale.

3) If the display becomes locked on 8888, this often indicates low batteries. However sometimes it also may indicate a serious zero mark error. This means when you turn the scale on, it can't determine what zero is (a slight zero mark error will cause situation #1 above) Thus, if new batteries do not fix this error the scale will have to be sent to us for replacement under our warranty program.

4) If the display shows UNST, this means the scale is not stable. Please try operating the scale on a more stable surface and be sure nothing is on the tray or stuck under the tray when you turn it on. If this situation persists, it may be an indication of fatal load cell damage. Try calibrating the scale - if this does not work please follow the warranty instructions. Although the AXE is designed to be extremely durable, It's important that you never overload or drop/shock the scale. Scales are delicate instruments and unlike Cellular phones, scales have delicate sensors that determine how much an item weighs. If you drop or shock your scale, these sensors "feel" the shock and are sometimes destroyed. This happens with all digital scales. We design our scales to be as resistant to shock or drops as possible, however there is no way for us to protect 100% against load cell or sensor damage. A welltreated scale will provide years of reliable and accurate weighing. However an abused scale will only work until it's sensors are damaged.

* UNIT (the left key) Unit (mode) selection: Press and hold the left key to change weight unit, You can select g(grams), oz(ounces), ozt(troy ounces), dwt(pennyweight), ct(carats), gn(grains) or lb(pounds).

* ON/OFF Press this key to turn unit on,Once the units is on, press and hold the same key for 5 seconds to turn the scale off.

* Tray Operation: The spinning locking mechanism of the AXE2 is designed to protect protect the scale from accidentel overload when it is closed. Simply spin the cover counter clockwise so that the LCD and buttons will be hidden from view. The scale is now ready for storage To operate the sale simply spin the cover clockwise until it is fully open (you'll feel it lock into place). Now the scale is ready for use. NOTE: Always close the cover when storing the scale. Please be sure the cover is fully open before operation.

* ZERO (TARE) After the scale is ON, you can press to reset the scale to zero. This can be used for eliminating from a sample (or a Tray/Container) the weight value of a container weight is permanently removed for the remainder of the procedure. Just turn the scale ON, place the tray or container on weighing platform, and press the key. The scale will show 0.0 (or 0.00) on the display and you can add items to the container/tray.

Digital scales are great for everyday use. I like to
buy digital scales or digital scale at cheap prices. I especially love the electronic
scales, electronic scale, kitchen scales, laboratory scales, laboratory scale,
kitchen scales, kitchen scale or even pocket scale or pocket scales. The best
kind of scale there is are the hand held scale or compact scale. Gram scales
are usually perfect for the hobbyist or the renown. If you need postal scales
or postal scale you can find them here.