IMPORTANT NOTE REGARDING
CALIBRATION /
SCALE MANUAL CORRECTION
In order to enter
calibration mode you must quickly press and release the zero and on/off keys.
If you hold the keys for too long the scale will simply turn back off. So, to
enter calibration mode press the Z key then press ON/OFF and quickly release the
keys. The scale will enter calibration mode and you can then follow the
regular calibration instructions.
Low Batteries & Faulty AC Adaptors are the #1 cause of scale malfunction and inaccuracy! We test all of our scale returns from consumers. Fully
60% of consumer returns are simply dead or low batteries. This sounds silly but it's true! A scale will perform slowly, or read inaccurately when it has low batteries. Please replace the batteries often (and only use good quality batteries). We include good quality batteries with all of our scales but batteries can run low in storage.
A Faulty AC adaptor can cause your scale to act unstable with numbers "jumping"
all around. Please test your scale with a good set of batteries (instead
of the AC adaptor) to determine if perhaps the AC adaptor is faulty.
Please click here for
the full printable 2004-2007 iBal 101 Manual
SAFETY PRECAUTIONS
All safety messages are identified by the words "WARNING"
and "CAUTION". These words mean the following:
|
eq \o\ac(△,!)!WARNING |
Important information to alert you to a
situation that might cause serious injury and damage to your property if
instructions are not followed. |
|
eq \o\ac(△,!)!CAUTION |
Important information that tells how to prevent
damage to the equipment. |
When using the i101, the following safety precautions
should always be followed.
eq \o\ac(△,!)!
WARNING
Use only the correct AC adaptor with the scale. Other
adaptors may cause permanent electrical damage.
eq \o\ac(△,!)!
CAUTION
Avoid placing the scale in
direct sunlight, this may cause discoloration or malfunction.
Replace all batteries at the
same time – Do not replace only a portion of the 6 batteries as this may cause a
malfunction.
If the scale is not to be used
for a long period of time, remove all batteries from the battery compartment to
avoid leakage, which may cause damage to this scale.
Avoid overloading the scale, as this may cause
permanent damage and void your warranty – do not EVER put more then 100 grams on
the scale.
Keep the scale away from water
– this scale is not water resistant. Shock, injury and electrical damage can
occur if used in a wet location!
Matter charged with static
electricity could affect accuracy. Discharge all static electricity. For
example, one method is to use Static-Guard spray, and spray it on both sides of
the weighing platform.
1
Counting procedures
1.Press [ON/OFF] to turn on
the scale.
Wait for "0" to appear on the
display.
2.Start the Count Procedure
If necessary, press [ZERO]
key to set the display to “0”.
3. Place a given number
of samples of an item on the pan (the Sample Size should be either 10、20、50
or 100 pieces). The weight of these samples will show on the display.
4. Press the [UNIT]
key several times to put the scale in PCS mode (the indicator should be on pcs).
5. Select the sample
size (the same as you chose above) by pressing the [PCS] key (press it as many
times as necessary to put it in the correct sample size (the sample size is the
same as in step three =10、20、50or100),
6. Press the [UNIT]
key, the display will show “pass”, then after 2 seconds, the scale will
remember the sample size you selected and show the starting sample size on the
display
(you can now remove the samples
if you want to return the scale to 0)
7. Place the items that
you want counted onto the tray and the total number of items will show on the
display.
8. Press the “UNIT” key
to exit the counting function and return to normal weighing or you can press
ON/OFF to turn the scale off .
NOTE: the weight of unit sample
> 10e.
FUNCTION SETTINGS
1. Enter
function setting mode.
The scale should be
OFF. Press and hold [UNIT] first, then press and release [ON/OFF]
(while keeping [UNIT] pressed) the display will show LIGHT or
A_OFF arrow and the display will show 888888 Now you can release the UNIT
key.
2. Selection
of auto backlight mode.
Press
[UNIT] to select auto backlight function, the display will show an arrow
next to LIGHT , this means auto backlight function active, press
[UNIT] again, the display will not show the arrow next to LIGHT, this
means the auto backlight function is inactive, i.e. shut off backlight function.
3. Selection
of auto shut off mode.
Press
[ZERO] to select auto shut off function, the display will show an arrow next
to A_OFF, this means auto shut off function active. Press [ZERO]
again, the display will not show an arrow next to A_OFF, this means the
auto shut off function is inactive.
4. Return to
weighing mode.
Press [ON/OFF] to turn the power off, press [ON/OFF] again to
turn on the power and the scale will return to normal weighing mode.
CALIBRATION
On some model i201 scales
there is a different calibration entry method The correction is that to enter
calibration mode you must quickly press and release the zero and on/off keys.
If you hold the keys for too long the scale will simply turn back off.
So, to enter calibration mode press the Z key then press on/off and quickly
release the keys. The scale will enter calibration mode and you can
then follow the regular calibration instructions.
When to calibrate - calibration is RARELY required.
Calibration may be required when the scale is first set up
for use, or if the scale is moved to a different altitude or gravitation. This
is necessary because the weight of a mass in one location is not necessarily the
same in another location. Also, with time and use, mechanical deviations can
occur.
How to calibrate
**you must have an
accurate 100 gram weight or combination of weights in order to calibrate**
1. The scale must be powered OFF. Press and hold [ZERO] first,
then Press [ON/OFF] while keeping the [ZERO] button depressed,
then release both keys. The display will show the A/D value (a series of random
numbers).
2.Wait
for the stable indicator to be displayed, Press [UNIT] the display will
show "00SAVE", then after 2 seconds the display will return to the random
A/D value.
3.Place 200 grams of weight(s) on platform. Wait for the stable
indicator to display, then press [UNIT], the display will show "00SAVE".
Calibration is complete, remove the weight(s).
4.Turn the scale OFF (Press the ON/OFF key) , then turn it back ON and
check some weight readings. If calibration is still incorrect, repeat
calibration but try it more slowly.
If calibration is repeatedly
incorrect then: 1) Replace the batteries, 2) Try a more stable surface or
vibration and interference free location, 3) Be sure there is no static
electricity on the scale (use an anti-static spray to remove static electricity)
FEATURES
l
Auto shut off (optional)
●
Auto calibration
l
Auto zero tracking
●
Auto backlight (optional)
l
Low battery indication
● g
/oz/mg/ozt/dwt/ct/gn/pcs conversion
l
Large LCD (6digits 15 mm high)
● with
counting function
l
Large square weighing tray
l
Net weight/stability indication
0.005 gram
division – professional accuracy
WEIGHING
Before weighing
Whenever possible, please allow the scale to warm up for 30
seconds after first turning the power on so that the scale will function
properly and accurately.
Error messages
EEEE: Overload
:
Low battery
Calibration may be required before weighing.
Calibration is rarely needed. Read "CALIBRATION" first and
if necessary, calibrate your scale for accurate weighing.
Weighing procedures
1.Press [ON/OFF] to turn on the scale.
When the power is turned on, all
display segments appear for a few seconds and "0" will appear on the display.
2.Select the weighing unit with [UNIT].
Press [UNIT] to select a
weighing unit "g"、"oz"
、"mg"、"ozt"
、"dwt"、"ct"、"gn"
or "pcs".
Once the unit has been
selected, the selected unit will be displayed next to the weight value.
3.Start weighing
If you do not use a container
for weighing,
Verify the reading is "0". If
not, press [ZERO] to display "0".
Place objects on the weighing platform to weigh.
When the reading becomes stable,
the stable indicator is displayed.
If you use a container for
weighing,
Place an empty container on the
platform.
Wait for the stability indicator to be displayed and press
[ZERO].
Place the objects to be weighed in the container.
When the reading becomes stable, the stable indicator is
displayed.
(trunacated)
|